HiTemp140-FP High Temperature Data Logger with a Flexible RTD Probe
The HiTemp140-FP is a durable, user friendly high temperature data logger featuring a long, flexible RTD probe with a narrow diameter, making it ideal for use in steam sterilization and lyophilization processes. Commonly used for mapping, validation and monitoring of high temperature surfaces and environments, this stainless steel data logger is available in several models. The flexible probe is coated with PFA insulation and can withstand temperatures up to 260°C. The body of the logger cannot exceed 140°C. The HiTemp140-FP probe design is narrow and lightweight making it ideal for placement within small vials, tubing, test tube and other small diameter or delicate applications.
Using the latest MadgeTech Software, starting, stopping and downloading the HiTemp140-FP is simple and easy. To use, simply place the HiTemp140-FP in the IFC400 or IFC406 docking station (sold separately). Using the software, an immediate or delay start can be chosen, as well as the reading interval. Start the data logger, remove it from the docking station and the device is ready to be deployed. Graphical, tabular and summary data is provided for analysis and data can be viewed in °C, °F, K or °R. The data can also be automatically exported to Excel® for further calculations.
Features
±0.1 °C (± 0.18 °F) Accuracy
Probe Measurement Range -200°C to +260°C
Operating Environment -40 °C to +140 °C
Submersible (IP68)
User-Replaceable Battery
Durable
Programmable Start and Stop Time
Five probe lengths: 6, 12, 24, 36 and 72 inches
Battery life indicator
HiTemp140-FP Benefits
Simple Setup and Installation
Minimal Long-Term Maintenance
Long-Term Field Deployment
Applications
Autoclave Verification and Mapping
Lyophilization
Monitoring High-Temperature Surfaces
Container Mapping
Measurements Inside Small Vials and Tubing
Specifications
TEMPERATURE
Temperature Sensor
Flexible RTD Probe
Probe Measurement Range
-200 °C to +260 °C (-328 °F to +500 °F)
BODY OF LOGGER CANNOT EXCEED 140 °C
Temperature Resolution
0.01 ºC (0.02 ºF)
Calibrated Accuracy
±0.1 °C from +20 °C to +140 °C
±0.3 °C from -20 °C to +19.99 °C
±0.4 °C from -40 °C to -20.01 °C
GENERAL
Start Modes
Software programmable immediate start
Delay start up to 18 months in advance
Stop Modes
Manual or Timed (specific date and time)
Data Logger Response Time
In Air
In Water
(hours : minutes : seconds : fractions of a second)
t60 – 00:00:30:00
t60 – 00:00:03:50
t90 – 00:01:10:00
t90 – 00:00:06:50
Trigger Settings
High and Low limits may be set. Once data meets or exceed sets limits, the device will record to memory. Bi-level start and stop triggers can also be programmed. Users can specify the number of readings to take after the device triggers
Memory
32,256 readings
Readings in Trigger Settings Mode
6,451 readings
Real Time Recording
May be used with PC to monitor and record data in real time
Password Protection
An optional password may be programmed into the device to restrict access to configuration options. Data may be read out without the password
Wrap Around
Yes
Reading Rate
4 readings per second up to 1 reading every 24 hours
Calibration
Digital calibration through software
Calibration Date
Automatically recorded within device
Battery Type
3.6V high-temperature lithium battery included; user replaceable
Battery Life
1 year typical (1 minute reading rate at 25 °C/77 °F)
Data Format
Date and time stamped °C, K, °F, °R, or mV
Time Accuracy
±1 minute/month at 25 °C (77 °F)
Extended Operation: ±20 minutes/month at 140 °C (±450 ppm)
Computer Interface
IFC400 OR IFC406 USB docking station and adapter required; 125,000 baud
Operating System Compatibility
Windows XP SP3 or later
Software Compatibility
MadgeTech Standard Software version 4.2.1.1 or later.
MadgeTech Secure Software version 4.2.0.1 or later.
Operating Environment
-40 °C to +140 °C (-40 °F to +284 °F)
0 %RH to 100 %RH, 0.002 PSIA to 100 PSIA
IP Rating
IP68
Dimensions (Body)
2.95 in x 0.97 in x 0.97 in
(75 mm x 24.6 mm x 24.6 mm)
Dimensions (Probe)
HiTemp140-FPST-6: 6 in x 0.13 in (152 mm x 3.2 mm)
HiTemp140-FPST-12: 12 in x 0.13 in (305 mm x 3.2mm)
HiTemp140-FPST-24: 24 in x 0.13 in (609 mm x 3.2mm)
HiTemp140-FPST-36: 36 in x 0.13 in (914 mm x 3.2mm)
HiTemp140-FPST-72: 72 in x 0.13 in (1829 mm x 3.2 mm)
Weight
3 oz (85 g)
Material
Body: 316 Stainless Steel, Radel
Probe: PFA Insulated Cable with Stainless Steel Tip
Approvals
CE
BATTERY WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT, CHARGE, FORCE OVER DISCHARGE, CRUSH, PENETRATE, OR INCINERATE. BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE 150 °C (302 °F).
Software
MadgeTech 4 Software
The MadgeTech 4 Software is designed to streamline the process of downloading and reviewing data in a user-friendly interface. The software offers many features and options, allowing for complete customization to satisfy regulatory requirements. MadgeTech 4 Software comes standard with any MadgeTech data logger and free updates are available to download from our website.
Reviews
There are no reviews yet.